RMIT University
Browse

Testing organic toxicants on biomicrofluidic devices: Why polymeric substrata can lead you into trouble

conference contribution
posted on 2024-10-31, 20:48 authored by Yushi Huang, Rhys Cartlidge, Feng Zhu, Dayanthi NugegodaDayanthi Nugegoda, Donald WlodkowicDonald Wlodkowic
Advances in microfabrication technologies and manufacturing over last decade, allowed for inexpensive prototyping of microfluidic chip-based devices for biomedical studies in biocompatible and optically transparent elastomeric polymers such as poly(dimethylsiloxane) (PDMS) and thermoplastics such as poly(methyl methacrylate) (PMMA). More resently, advanced additive manufacturing technologies such as stereolithography (SLA), capable of reproducing feature sizes less than 50 μm, pave a way towards a new generation of microfabrication techniques. The latter promise new methods to enable accelerated design, validation and optimisation of optical-grade biomicrofluidic Lab-on-a-Chip (LOC) devices. The main limitation, however, of virtually all polymers that are used to both manufacture LOC devices as well as to provide fluidic interconnects is their significant hydrophobicity. Conventionally the hydrophobic properties were postulated to impede wetting and priming of the polymeric chip-based devices and tubing interconnects. Such issues were often solved with plasma treatment or ethanol priming to help wet the polymeric substrata and also reduce the nucleation and persistence of air bubbles. In this work, we present evidence that use of certain hydrophobic polymers is a significant impediment in performing ecotoxicity tests of organic chemicals on biomicrofluidic devices. We report on electrostatic interaction between polymers and toxicants that lead to non-covalent adsorption and rapid depletion of chemicals from the tested media. This introduces a significant bioanalytical bias irrespectively of the fact that microfluidic tests are preformed under continuous perfusion.

Funding

Ecotoxicology-on-a-chip: towards smart devices in environmental biomonitoring

Australian Research Council

Find out more...

History

Start page

1

End page

7

Total pages

7

Outlet

Proceedings SPIE 9668, Micro+Nano Materials, Devices, and Systems

Editors

Benjamin J. Eggleton and Stefano Palomba

Name of conference

SPIE Micro+ Nano Materials, Devices, and Applications

Publisher

International Society for Optics and Photonics

Place published

Sydney, Australia

Start date

2015-12-06

End date

2015-12-06

Language

English

Copyright

© 2015 SPIE

Former Identifier

2006071647

Esploro creation date

2020-06-22

Fedora creation date

2017-03-21

Usage metrics

    Scholarly Works

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC