There is a need for simple and inexpensive methods to quantify potentially harmful persistent pesticides often found in our water-ways and water distribution systems. This paper presents a simple, relatively inexpensive method for the detection of a group of commonly used pesticides (atrazine, simazine and hexazinone) in natural waters using large-volume direct injection high performance liquid chromatography (HPLC) utilizing a monolithic column and a single wavelength ultraviolet¿visible light (UV¿vis) detector. The best results for this system were obtained with a mobile phase made up of acetonitrile and water in a 30:70 ratio, a flow rate of 2.0 mL min-1, and a detector wavelength of 230 nm. Using this method, we achieved retention times of less than three minutes, and detection limits of 5.7 ?g L-1 for atrazine, 4.7 ?g L-1 for simazine and 4.0 ?g L-1 for hexazinone. The performance of this method was validated with an inter-laboratory trial against a National Association of Testing Authorities (NATA) accredited liquid chromatography¿mass spectrometry/mass spectrometry (LC¿MS/MS) method commonly used in commercial laboratories.
NOTICE: this is the author’s version of a work that was accepted for publication in Talanta. Changes resulting from the publishing process, such as peer review, editing, corrections, structural formatting, and other quality control mechanisms may not be reflected in this document. Changes may have been made to this work since it was submitted for publication. A definitive version was subsequently published in Talanta, Vol 82, No 2, (2010) http://dx.doi.org/10.1016/j.talanta.2010.05.030