RMIT University
Browse

Evaluation of patient doses for routine digital radiography procedures toward establishing an institutional diagnostic reference levels: A case study in Sri Lanka

journal contribution
posted on 2024-11-02, 22:25 authored by Sachith Welarathna, Sivakumar Velautham, Mihira Wanninayake, Siva SarasanandarajahSiva Sarasanandarajah
The present study was conducted as part of a comprehensive work to establish National Diagnostic Reference Levels (NDRLs) in Sri Lanka for the first time. DRLs can be used as an effective optimization tool for identifying unusually high or low patient doses during X-ray examinations. This study aims to propose institutional DRLs (IDRLs) by measuring the kerma-area product (KAP) of adult patients undergoing routine projection X-ray examinations. The median and the 75th percentile KAP values obtained were compared with that of the single institution KAP values reported from India and Greece. This descriptive cross-sectional study was conducted in a public hospital in Uva province, Sri Lanka, with 400 adult patients aged 18–87 years and weighing 58 ± 20 kg. The patient-specific information (age, sex, weight, and height) and corresponding exposure parameters (tube voltage and current-exposure time product) were obtained. The KAP values were measured, and descriptive statistics were utilized for data analysis. The median KAP values obtained were proposed as IDRLs. The IDRLs in Gy.cm2 were 0.23 for cervical spine anterior–posterior (AP), 0.19 for cervical spine lateral (LAT), 0.10 for chest posterior–anterior (PA), 0.06 for knee joint AP, 0.05 for knee joint LAT, 1.47 for KUB AP, 0.85 for lumbar spine AP, 1.97 for lumbar spine LAT, 0.29 for shoulder joint AP, 0.61 for skull PA, and 0.60 for skull LAT examinations. The maximum to minimum ratio of KAP values ranged from 2.4 for KUB AP to 6.3 for the cervical spine AP examinations. The median and the 75th percentile of most of the examinations were comparable to corresponding KAP values reported by the countries mentioned above, except for the skull PA and LAT examinations. Accordingly, interquartile ranges of exposure parameters are recommended for skull examinations to improve the optimization of patient doses.

History

Related Materials

  1. 1.
    DOI - Is published in 10.1002/acm2.13852
  2. 2.
    ISSN - Is published in 15269914

Journal

Journal of Applied Clinical Medical Physics

Volume

23

Number

e13852

Issue

12

Start page

1

End page

11

Total pages

11

Publisher

John Wiley & Sons

Place published

United States

Language

English

Copyright

© 2022 The Authors. This is an open access article under the terms of the Creative Commons Attribution 4.0 International (CC BY 4.0) License

Former Identifier

2006120102

Esploro creation date

2023-03-25

Usage metrics

    Scholarly Works

    Licence

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC