Measuring dose in lung identifies peripheral tumour dose inaccuracy in SBRT audit

journal contribution
posted on 2024-11-03, 10:07 authored by Maddison Shaw, Jessica LyeJessica Lye, Andrew Alves, Joerg Lehmann, Masoumeh Sanagou, Moshi GesoMoshi Geso, Rhonda Brown
Purpose: Stereotactic Body Radiotherapy (SBRT) for lung tumours has become a mainstay of clinical practice worldwide. Measurements in anthropomorphic phantoms enable verification of patient dose in clinically realistic scenarios. Correction factors for reporting dose to the tissue equivalent materials in a lung phantom are presented in the context of a national dosimetry audit for SBRT. Analysis of dosimetry audit results is performed showing inaccuracies of common dose calculation algorithms in soft tissue lung target, inhale lung material and at tissue interfaces. Methods: Monte Carlo based simulation of correction factors for detectors in non-water tissue was performed for the soft tissue lung target and inhale lung materials of a modified CIRS SBRT thorax phantom. The corrections were determined for Gafchromic EBT3 Film and PTW 60019 microDiamond detectors used for measurements of 168 SBRT lung plans in an end-to-end dosimetry audit. Corrections were derived for dose to medium (Dm,m) and dose to water (Dw,w) scenarios. Results: Correction factors were up to −3.4% and 9.2% for in field and out of field lung respectively. Overall, application of the correction factors improved the measurement-to-plan dose discrepancy. For the soft tissue lung target, agreement between planned and measured dose was within average of 3% for both film and microDiamond measurements. Conclusions: The correction factors developed for this work are provided for clinical users to apply to commissioning measurements using a commercially available thorax phantom where inhomogeneity is present. The end-to-end dosimetry audit demonstrates dose calculation algorithms can underestimate dose at lung tumour/lung tissue interfaces by an average of 2–5%.

History

Related Materials

  1. 1.
    DOI - Is published in 10.1016/j.ejmp.2023.102632
  2. 2.
    ISSN - Is published in 11201797

Journal

Physica Medica

Volume

112

Number

102632

Start page

1

End page

9

Total pages

9

Publisher

Elsevier

Place published

United Kingdom

Language

English

Copyright

Crown Copyright © 2023 Published by Elsevier Ltd on behalf of Associazione Italiana di Fisica Medica e Sanitaria. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).

Former Identifier

2006124391

Esploro creation date

2023-08-09