Medical physics aspects of the synchrotron radiation therapies: Microbeam radiation therapy (MRT) and synchrotron stereotactic radiotherapy (SSRT)

journal contribution
posted on 2024-11-01, 22:31 authored by Elke Brauer-Krisch, Jean-Francois Adam, Enver Alagoz, Stefan Bartzsch, Jeffrey Crosbie, Carlos DeWagter, Andrew Dipuglia, Mattia Donzelli, Simon Doran, Pauline Fournier, John Kalef-Ezra, Angela Kock, Michael Lerch, Ciara McErlean, Uwe Oelfke, Pawel Olko, Marco Petasecca, Marco Povoli, Anatoly Rosenfeld, Erik Siegbahn, Dan Sporea, Bjarne Stugu
Stereotactic Synchrotron Radiotherapy (SSRT) and Microbeam Radiation Therapy (MRT) are both novel approaches to treat brain tumor and potentially other tumors using synchrotron radiation. Although the techniques differ by their principles, SSRT and MRT share certain common aspects with the possibility of combining their advantages in the future. For MRT, the technique uses highly collimated, quasi-parallel arrays of X-ray microbeams between 50 and 600keV. Important features of highly brilliant Synchrotron sources are a very small beam divergence and an extremely high dose rate. The minimal beam divergence allows the insertion of so called Multi Slit Collimators (MSC) to produce spatially fractionated beams of typically ~25-75 micron-wide microplanar beams separated by wider (100-400 microns center-to-center(ctc)) spaces with a very sharp penumbra. Peak entrance doses of several hundreds of Gy are extremely well tolerated by normal tissues and at the same time provide a higher therapeutic index for various tumor models in rodents. The hypothesis of a selective radio-vulnerability of the tumor vasculature versus normal blood vessels by MRT was recently more solidified.SSRT (Synchrotron Stereotactic Radiotherapy) is based on a local drug uptake of high-Z elements in tumors followed by stereotactic irradiation with 80keV photons to enhance the dose deposition only within the tumor. With SSRT already in its clinical trial stage at the ESRF, most medical physics problems are already solved and the implemented solutions are briefly described, while the medical physics aspects in MRT will be discussed in more detail in this paper.

History

Related Materials

  1. 1.
    DOI - Is published in 10.1016/j.ejmp.2015.04.016
  2. 2.
    ISSN - Is published in 11201797

Journal

Physica Medica: European journal of medical physics

Volume

31

Start page

568

End page

583

Total pages

16

Publisher

Elsevier Ltd

Place published

United Kingdom

Language

English

Copyright

© 2015 Associazione Italiana di Fisica Medica

Former Identifier

2006053981

Esploro creation date

2020-06-22

Fedora creation date

2015-07-06

Usage metrics

    Scholarly Works

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC