The aim of this study is to present an efficient method to generate imager-specific Monte Carlo (MC)-based dose kernels for amorphous silicon-based electronic portal image device dose prediction and determine the effective backscattering thicknesses for such imagers. EPID field size-dependent responses were measured for five matched Varian accelerators from three institutions with 6 MV beams at the source to detector distance (SDD) of 105 cm. For two imagers, measurements were made with and without the imager mounted on the robotic supporting arm. Monoenergetic energy deposition kernels with 0¿2.5 cm of water backscattering thicknesses were simultaneously computed by MC to a high precision. For each imager, the backscattering thickness
required to match measured field size responses was determined. The monoenergetic kernel method
was validated by comparing measured and predicted field size responses at 150 cm SDD, 10x10 cm2 multileaf collimator (MLC) sliding window fields created with 5, 10, 20, and 50 mm gaps, and a head-and-neck (H and N) intensity modulated radiation therapy (IMRT) patient field.