Monte Carlo verification of gel dosimetry measurements for stereotactic radiotherapy
journal contribution
posted on 2024-11-01, 10:59authored byTanya Kairn, Michael Taylor, B Crowe, Leon Dunn, Rick FranichRick Franich, J Kenny, R Knight, V Trapp
The quality assurance of stereotactic radiotherapy and radiosurgery treatments requires the use of small-field dose measurements that can be experimentally challenging. This study used Monte Carlo simulations to establish that PAGAT dosimetry gel can be used to provide accurate, high-resolution, three-dimensional dose measurements of stereotactic radiotherapy fields. A small cylindrical container (4 cm height, 4.2 cm diameter) was filled with PAGAT gel, placed in the parietal region inside a CIRS head phantom and irradiated with a 12-field stereotactic radiotherapy plan.