RMIT University
Browse

Synchrotron Microbeam Radiotherapy In A Commercially Available Treatment Planning System

journal contribution
posted on 2024-11-02, 11:08 authored by Christopher Poole, Liam Day, Peter Rogers, Jeffrey Crosbie
In this paper we use a research licensed version of the Eclipse (TM). treatment planning system (TPS) from Varian Medical Systems, Inc. to develop a novel treatment planning environment for synchrotron microbeam radiotherapy (MRT). This research license allows for customised dose calculation algorithms to be integrated with the clinical work-flows in Eclipse that are typical to modern radiotherapy. In this way, all the functionality of the Eclipse TPS familiar to many radiation oncology professionals is retained for application to MRT. Our TPS is designed for the dynamic MRT modality that has been developed for the imaging and medical beamline (IMBL) at the Australian synchrotron. The TPS uses a very simple algorithm for primary dose calculation, and we estimate the peak to valley dose ratio (PVDR) and hence the valley dose from data derived from a separate Monte Carlo simulation. At this stage the algorithm is a fast approximation with only simple radiological path length corrections to account for tissue inhomogeneities. However, the algorithm can highlight to oncology clinicians the strengths and limitations of synchrotron MRT, and furthermore, serves as a proof of concept and a start point for implementingmore advanced dose calculation algorithms. For the treatment itself, the incident synchrotron broad beam is collimated to a30mm wide by 1mm high field which illuminates the MRT collimator, whichin turn produces 50 mu m wide vertical microbeams separated at 00 mu m centre-to-centre. The sample and amask is then dynamically swept through this array of microbeams, producing a dose of radiation in the sample that is conformal to a patient specific mask aperture. All properties of this irradiation methodology can be configuredwithin the TPS and used for the design of MRT treatments. We present implementation details of our TPS for MRT nd describe the hyperbolic relationship between PVDR and depth, and use it for calculating the water equivalent PVDR at different radiological depths.

History

Related Materials

  1. 1.
    DOI - Is published in 10.1088/2057-1976/aa5f1a
  2. 2.
    ISSN - Is published in 20571976

Journal

Biomedical Physics & Engineering Express

Volume

3

Number

025001

Issue

4

Start page

1

End page

13

Total pages

13

Publisher

Institute of Physics

Place published

United Kingdom

Language

English

Copyright

© 2017 IOP Publishing Ltd

Former Identifier

2006095856

Esploro creation date

2020-09-08

Usage metrics

    Scholarly Works

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC