posted on 2024-11-23, 16:19 authored by Thu Tran
This project involved designing and building a radiochromic film (RCF) densitometer using a still digital camera as the light detector and light emitting diode, LED, as the light source. The behaviour of the LED and charged coupled device (CCD) in the still digital camera, under different light exposure settings (by changing LED current and camera shutter speed) were measured and an optimal setting was determined. Additionally, methods were devised and tests were carried out in order to spread the illumination area of the single light source. Uniform spreading of the LED illumination area was possible by the use of two diffusers placed at an optimum separation distance that was determined in this work.

The usefulness of this custom-made RCF densitometer was demonstrated by using this device to image exposed RCF and using the film analysis software, Image J, to determine the film absorbed dose. Two clinical situations were examined: open and virtual wedge radiation beams.

It was concluded that still digital cameras can be used in RCF densitometers provided they can capture and store raw images, a single diffused LED can illumination an area large enough for RCF densitometry and appropriate film analysis software is needed to extract and handle the large volume of greyscale data from the RCF.

History

Degree Type

Masters by Research

Imprint Date

2008-01-01

School name

School of Science, RMIT University

Former Identifier

9921863640401341

Open access

  • Yes

Usage metrics

    Theses

    Categories

    No categories selected

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC